374
I
NTRODUCTION
Bacterial contamination of treated water is a matter of
serious concern in hemodialysis (HD)
centers and im-
provement of microbiological quality is a permanent
challenge for professionals at these units.
Patients on regular HD treatment are exposed to hun-
dreds of liters of dialysis water every week through the
extracorporeal circulation (1). Pyrogenic reactions
and bacteremia are potential consequences of poor
microbiological quality of treated water for the prepa-
ration of dialysate and reprocess in HD (2)
.
Several factors favour bacterial growth in the water
treatment systems for HD, including sand and carbon
filters and softeners, or parts with low circulation such
as tanks and taps (3). Formation of biofilms that facili-
tate bacterial persistence at different points of the sy-
stem and protect bacteria from disinfection
also in-
creases the risk of contamination and high endotoxin
levels in water (4).
Despite continuous efforts at these centers to improve
the microbiological quality of treated water for HD,
water microbiological quality parameters must be con-
stantly assessed in order to avoid undesirable conse-
quences and implement improvement programs.
We report the results of a three-year study of the bac-
teriological quality of water samples in 32 out of a total
of 38 HD centers throughout Uruguay. All the sam-
ples were processed at our private independent labo-
ratory, as part of the centers' quality assurance pro-
gramme. The results had to be reported annually to a
special commitee, acting on behalf of the health
authorities. For this study, samples were obtained at
different points of the purification process and exami-
ned for heterotrophic bacteria, total coliforms, and
Pseudomonas aeruginosa
.
J N
EPHROL 2002; 15: 374-379
www.sin-italia.org/jnonline/vol15n4/
Microbiological quality of hemodialysis
water in a three-year multicenter study
in Uruguay
Pablo Zunino, Lucía Beltrán, Laura Zunino, Hernán Méndez,
Verónica Percovich, Rossana Rocca, Beatriz
Antonelli
Beltrán - Zunino Laboratory (B-Z SRL), Montevideo - Uruguay
ABSTRACT:
Background:
The microbiological quality of treated water is a very important
issue in hemodialysis (HD)
centers. Water treatment may have a dramatic effect on microbial
contamination because of bacterial colonisation of
the different parts of the system such as storage tanks, softeners or
deionisers. Therefore, HD centers must have
stringent quality programmes including regular water monitoring for
microbiological analysis. We report the results
of a three-year study (July 97 to June 2000) including
bacteriological quality analysis of 3129 water samples from 32
out of 38 HD centers throughout Uruguay.
Methods:
Bacteriological analysis of water samples was based on heterotrophic
count, total coliform count and
Pseudomonas aeruginosa
presence/absence, according to the procedures proposed by the
Association for the Ad-
vancement of Medical Instrumentation (AAMI).
Results:
Heterotrophic counts of 83% of the samples were under 200 colony
forming units (CFU)/mL (AAMI limit of
compliance) and the water samples after the final bactericidal
treatment showed 99% compliance. The points show-
ing the worst results were softeners and deionisers (60% acceptance).
Conclusions:
In comparison with a similar study in Uruguay, from January 96 to
June 97 at the same microbiological
laboratory, there has been a marked improvement in the
microbiological quality of water for hemodialysis.
Key words:
Water, Hemodialysis,
Microbiological quality, Uruguay

Zunino et al
375
S
UBJECTS AND METHODS
Water sampling.
Water was sampled monthly at most of
the centers, as recommended by the Association for
the Advancement of Medical Instrumentation (AA-
MI) (5) over a three-year period in 32 centres from va-
rious regions in Uruguay (July 1997 to June 2000).
Sampling was done either by unit workers or by labo-
ratory workers, depending on the dialysis center.
Water was sampled aseptically according to the Ameri-
can Public Health Association (APHA) methods (6).
About 120 mL of water were collected in commercial
sterile polypropylene bottles (Bioster
) after 1 to 2 mi-
nutes of free flow through the faucets.
Water sampling sites were as follows: sand filter, activa-
ted carbon filter, softener or deioniser, reverse osmo-
sis, treated water storage tank, HD room, membrane-
washing room (where dialysis membranes are washed
and treated for reuse), recirculation loop, and post re-
verse osmosis bactericidal treatment area (UV light or
ozone). In all HD centers, water
samples obtained
from the HD room faucet or from the nearest availa-
ble sampling point were considered as treated.
In order to compare the performance of individual
HD centers, only those with at least 12 treated water
samples during the whole period and at least four in
the last year were considered.
A total of 3129 water samples were analysed during the
three years of the study.
Bacteriological analysis.
All bacteriological tests were do-
ne according to the Standard Methods for the Exami-
nation of Water and Wastewater (6). All bacteriologi-
cal media were from Difco, Detroit, Mi. (USA).
Heterotrophic bacteria counts were done in triplicate
on Plate Count Agar by the spread plate technique
and plates were incubated for 72 h at 35° C. Total co-
liforms were determined by membrane filtration: 100
mL of water was filtered using 0.45 µm pore size mem-
branes (Pall Gelman Laboratory). Membranes were
placed on m-Endo plates and incubated for 24 h at 35º
C. Coliforms were detected by production of typical
red colonies with a metallic surface sheen or atypical
dark red colonies without sheen. All were confirmed
for gas production in lauryl tryptose broth tubes, in-
cubated at 35º C for 48 h.
Pseudomonas aeurginosa
was also assessed: 10 mL of wa-
ter were added to asparagine broth and incubated at
35ºC for 48 h. Tubes showing turbidity and produc-
tion of fluorescent pigment, assessed with a long-wave-
length UV light lamp, were subcultured in Cetrimide
Milk Agar for determination of
P. aeruginosa
(6).
Bacteriological acceptance limits.
Standards for microbio-
logical water quality recommended by the AAMI were
used to determine compliance of water samples, ac-
cording to national health authority criteria (5). The-
se guidelines establish an acceptable limit of 200 co-
lony-forming units (CFU)/mL of heterotrophic bacte-
ria for treated water used for HD and 2000 CFU/mL
for dialysate.
R
ESULTS
Heterotrophic bacteria counts from all sampling sites.
Ac-
ceptable heterotrophic bacteria counts (<200
CFU/mL) of all samples varied from 60%, in the sof-
tener or deioniser samples, to 99% after bactericidal
treatment (UV light or ozone) (Tab.
I).
Heterotrophic bacteria counts from samples of treated water.
When heterotrophic bacteria counts in treated water
were considered during the three-year period, 83%
of the samples were below 200 CFU/mL, 9%
between 200 and 1000 CFU/mL and 8% were over
1000 CFU/mL. When results were analysed for each
year of the study, an improvement in bacteriological
quality was observed since the percentage of samples
with counts below 200 CFU/mL was 73 in the first
year, 87 in the second and 89 in the third.
Heterotrophic bacteria counts of samples from HD room fau-
cets.
Heterotrophic bacteria counts of samples from
HD room faucets improved with time. The percenta-
ge of acceptable samples in the first year was 58, ri-
sing to 78 in the second, and 87 in the third and final
year.
Heterotrophic bacteria counts related to sampling month.
Geometric means of heterotrophic bacteria counts
in samples from HD room faucets were analysed in
relation to the collection month in order to assess
the influence of ambient temperature. During the
winter months (June to September) counts were no-
tably lower than over the rest of the year, when am-
bient temperature is higher (Fig. 1). The correlation
between the geometric mean of heterotrophic bacte-
ria counts and the mean temperature was 0.7, confir-
ming that temperature has an important impact on
microbiological water quality.
Fig. 1 - Geometric mean of heterotrophic counts (CFU/mL) at
hemodialysis room and ambient temperature
(ºC) related to
sampling month, expressed as means for the three
years.
CFU/mL (geometric mean)
Temperatur
e (°C)
Month
Hemodialysis
water in Uruguay
376
Heterotrophic bacteria counts of treated water samples in
individual HD centers.
Microbiological quality at dif-
ferent HD centers varied widely in treated water.
Samples showing acceptable counts ranged from
100% to a minimum of 32.8% (Tab. II). However,
water microbiological quality improved over the
study period. When results for only the last year
were assessed, the worst center had more than 60%
of acceptable samples (Tab. III).
Pseudomonas aeruginosa in water at all sampling sites. P.
aeuruginosa
was recovered from water samples from
all sampling points, except after reverse osmosis bac-
tericidal treatment (UV light or ozone),
immediately
after the treated water storage tank, and from the
membrane-washing room. At the other sampling
points there was a low percentage of positive sam-
TABLE I -
HETEROTROPHIC BACTERIA COUNTS AT ALL SAMPLING POINTS
Sampling point
Number of samples
<200
a
%
200-1000
b
%
>1000
c
%
Sand filter
58
98
0
2
Activated carbon filter
596
65
20
15
Softener or deionizer
527
60
21
19
Reverse osmosis
441
88
9
3
Treated water storage tank
291
90
5
5
HD room
696
76
14
10
Membrane washing room
123
89
8
3
Recirculation loop
305
88
5
7
After bactericidal treatment
92
99
0
1
Total
3129
a - Percentage of samples with heterotrophic bacteria counts below
200 colony forming units (CFU)/mL
b - Percentage of samples with heterotrophic bacteria counts between
200 and 1000 CFU/mL
c - Percentage of samples with heterotrophic bacteria counts over
1000 CFU/mL
TABLE II -
HETEROTROPHIC BACTERIA COUNTS IN INDIVIDUAL HEMODIALYSIS
(HD) CENTERS (JULY 1997 JUNE 2000)
HD Center
<200
a
%
200-1000
b
%
>1000
c
%
N
d
1
100
0
0
36
2
98.2
1.8
0
56
3
97.6
0
2.4
41
4
95.6
4.4
0
45
5
95.2
2.4
2.4
42
6
93.6
2.1
4.3
47
7
93.3
3.3
3.3
30
8
90
4
6
50
9
88.5
7.7
3.8
26
10
80
20
0
15
11
80
15
5
40
12
78.6
21.4
0
14
13
76.9
23.1
0
52
14
75
25
0
28
15
74.1
14.8
11.1
27
16
73.4
21.9
4.7
64
17
72.1
19.8
8.1
86
18
68.3
17.1
14.6
41
19
56.7
20
23.3
30
20
55.6
2.2
42.2
45
21
32.8
35.9
31.3
64
Total
879
a - Percentage of samples with heterotrophic bacteria counts below
200 colony forming units (CFU)/mL
b - Percentage of samples with heterotrophic bacteria counts between
200 and 1000 CFU/mL
c - Percentage of samples with heterotrophic bacteria counts over
1000 colony forming units CFU/mL
d - Number of samples
Zunino et al
377
ples, ranging from 0.4% (one sample out of 269), at
the treated water storage tank sampling point, to
2.7% (17/632) in the HD room (Tab. IV).
Total coliforms in water from all sampling sites.
Co-
liforms were recovered from water at five sampling
points: activated carbon filter, softener or deioniser,
reverse osmosis, treated water storage tank, and HD
room. Coliforms were not recovered from the sand
filter, membrane washing room, recirculation loop,
or after bactericidal treatment (UV light or ozone)
(Tab. V). The percentage of positive samples was hi-
ghest after the carbon filter (0.8%, 4/511) and lowe-
st from the treated water storage tank (0.4%, 1/269)
(Tab. V).
TABLE III -
TOTAL HETEROTROPHIC BACTERIA COUNTS IN INDIVIDUAL HEMODIALYSIS
(HD) CENTERS (JULY
1999 JUNE 2000)
HD Center
<200
a
%
200-1000
b
%
>1000
c
%
N
d
1
100
0
0
10
2
100
0
0
25
3
100
0
0
13
4
92.3
7.7
0
13
5
90.9
9.1
0
11
6
100
0
0
14
7
75
12.5
12.5
8
8
100
0
0
18
9
100
0
0
9
10
81.8
18.2
0
11
11
64.3
21.4
14.3
14
12
78.6
21.4
0
14
13
85.7
14.3
0
14
14
80
20
0
10
15
75
25
0
4
16
100
0
0
16
17
67.7
19.4
12.9
31
18
100
0
0
10
19
100
0
0
4
20
100
0
0
9
21
75
25
0
20
22
100
0
0
6
23
85.7
14.3
0
7
Total
291
a - Percentage of samples with heterotrophic bacteria counts below
200 colony forming units (CFU)/mL
b - Percentage of samples with heterotrophic bacteria counts between
200 and 1000 CFU/mL
c - Percentage of samples with heterotrophic bacteria counts over
1000 CFU/mL
d - Number of samples
TABLE IV -
PSEUDOMONAS AERUGINOSA
IN SAMPLES OBTAINED AT DIFFERENT POINTS OF THE HD SYSTEM
Sampling point
Total samples
Positive samples
a
Positive
b
%
Sand filter
91
1
1.1
Activated carbon filter
511
10
2.0
Softener or deionizer
499
7
1.4
Reverse osmosis
432
8
1.8
Treated water storage tank
269
1
0.4
HD room
632
17
2.7
Membrane washing room
210
0
0
Recirculation loop
281
4
1.4
After bactericidal treatment
60
0
0
Total
2985
48
1.6
a - Number of samples containing
P. aeruginosa
b - Percentage of samples containing
P. aeruginosa
Hemodialysis
water in Uruguay
378
D
ISCUSSION
This three-year study assessed the bacterial quality of
water used in 32 out of a total of 38 HD centers throu-
ghout Uruguay. When heterotrophic bacteria counts
were related to the different sampling points, the hi-
ghest percentage of acceptable samples (<200
CFU/mL) was obtained after UV or ozone
bacterici-
dal treatment (99%). This suggests that the use of bac-
tericidal treatment devices after reverse osmosis units
in HD water purification systems has a remarkably be-
neficial effect on water bacterial quality. Water obtai-
ned after sand filters also gave a high percentage of ac-
ceptable samples (98%), suggesting that these filters
may not be critical for water bacterial quality. Howe-
ver, the number of samples obtained after UV or bac-
tericidal treatment and after the sand filter was lower
than the number of samples at other points of the sy-
stem. The lowest percentage of acceptable samples
was at the activated carbon filters (65%), softeners
and deionisers (60%) and HD rooms (76%). These re-
sults are in accordance with other reports that carbon
filters and softeners and deionisers are important risk
points since they can host bacterial growth. In addi-
tion, water going through them is already chlorine
free (7, 8).
Poor results in the HD rooms could indicate bad hy-
giene of room faucets since recirculating water quality
was notably better. However, when results for HD
room faucet samples were considered year by year, the
percentage of acceptable values rose encouragingly,
from 58% for the first year to 87% for the third. This
improvement might be due to the progressive applica-
tion of faucet disinfecting processes as routine at the
different HD centers.
Microbiological counts of treated water over the three
years showed wide variability in the different centers,
ranging from 100% (only one centre) to 32.8% of
acceptable samples. This indicates that HD centers in
Uruguay vary widely and that good practice and
methods need to be extended to all the centers.
However, looking at these measurements only for the
final year of the study, the situation had dramatically
improved. Ten centers out of 23 had 100% acceptable
samples (below 200 CFU/mL) and the worst one
showed 67.7% of acceptable samples. These results in-
dicate that, in general, control programmes have im-
proved and have had a beneficial effect on water bac-
terial quality.
The higher heterotrophic counts obtained during
warm weather suggest the need for maximum precau-
tions during this period when bacterial multiplication
is fastest. Laurence et al reported a similar correlation
between counts and warm seasons during a seven-year
multicenter study in Canada (9).
We also evaluated
Pseudomonas aeruginosa
and total co-
liforms in samples from different points of the treat-
ment systems.
P. aeruginosa
was recovered from 1.6%
of total samples. The highest percentage of positive
samples was in the HD room (2.7%), carbon filter
(2.0%), and at reverse osmosis (1.8%).
P. aeruginosa
was not detected in water from the membrane wa-
shing rooms or water obtained after UV or ozone
treatment.
Total coliforms were observed in 0.5% of samples.
The highest percentage was at the activated carbon fil-
ter (0.8%), reverse osmosis (0.7%), and softener or
deioniser and HD rooms (0.6%). Total coliforms and
Pseudomonas
recoveries were lower than those repor-
ted by Arvanitidou et al in a similar study from Greece
in 1997 (1).
In a previous survey in Uruguay, from January 1996 to
June 1997, 912 water samples from different points of
water treatment systems at 21 HD centers were analy-
sed to check the microbiological quality of water used
for HD and the efficiency of treatment (10). Only
TABLE V -
COLIFORMS IN SAMPLES OBTAINED AT DIFFERENT POINTS OF THE HD SYSTEM
Sampling point
Total samples
Positive samples
a
Positive
b
%
Sand filter
91
0
0
Activated carbon filter
511
4
0.8
Softener or deionizer
499
3
0.6
Reverse osmosis
432
3
0.7
Treated water storage tank
269
1
0.4
HD room
632
4
0.6
Membrane washing room
210
0
0
Recirculation loop
281
0
0
After bactericidal treatment
60
0
0
Total
2985
15
0.5
a - Number of samples containing total coliforms
b - Percentage of samples containing total coliforms
Zunino et al
379
63.4% of the samples were under the AAMI acceptan-
ce limit. The sampling point with the highest percen-
tage of acceptable values was reverse osmosis (85.3%)
and the one with the lowest percentage was the HD
room faucets (18%) (10).
The improvement seen in the findings of our three-
year study was probably due to the adoption of better
equipment in the treatment systems at most HD cen-
ters, the addition of bactericidal treatment devices af-
ter reverse osmosis and the implementation of more
efficient water quality assurance programmes inclu-
ding regular monthly sampling and thorough disin-
fection of HD room faucets. The microbiological qua-
lity of water used for HD after the present study can be
considered among the best results reported after simi-
lar studies (1, 9, 11).
Address for correspondence:
Lucía Beltrán, M.D.
Juan Paullier 1070
11200, Montevideo, Uruguay
bzmicro@adinet.com.uy
R
EFERENCES
1. Arvanitidou M, Spaia S, Katsinas C, Pangidis P, Constantinidis
T, Katsouyannopoulus V, Vayonas G. Microbial quality of wa-
ter and dialysate in all haemodialysis centres of Greece.
Nephrol Dial Transplant 1998; 13: 949-54.
2. Favero MS, Alter MJ, Bland LA. Dialysisassociated infections.
In Bennett JV, Brachman PS, Sanford JP (eds.), Hospital In-
fections. Boston: Little Brown, 1992; 375-403.
3. Du Molin GC, Coleman EC, Hedley-Whyte J. Bacterial colo-
nization and endotoxin content of a new renal dialysis water
system composed of acrylonitrile butadiene styrene. Appl
Env Microbiol 1987; 53: 1322-6.
4. Morin P. Identification of the bacteriological contamination
of a water treatment line used for haemodialysis and its disin-
fection. J Hosp Infect 2000; 45: 218-24.
5. Association for the Advancement of Medical Instrumentation
(AAMI). Water quality for dialysis, 3rd. Ed., AAMI WQD, Ar-
lington, 1998.
6. American Public Health Association (APHA). Standard Meth-
ods for the Examination of Water and Wastewater, APHA-
AWWA-WEF, 20th Ed.,1998; 9-33.
7. Russell AD, Hugo WB, Ayliffe GA (eds.). Disinfection,
Preser-
vation and Sterilization, Blackwell Scientific Publications
1992; 286-8.
8. Lauer J, Streifel A, Kjellstrand CM, De Roos R. The bacterial
quality of hemodialysis solution as
related to several environ-
mental factors. Nephron 1975; 15: 87-97.
9. Laurence RA, Lapierre ST. Quality of haemodialysis water: a
7-year multicenter study. Am J Kidney Dis 1995; 25: 738-50.
10. Beltrán L, Zunino L, Méndez H, Martínez I, Ramos J, Grosso
R, Armand Ugón D. Contaminación microbiana de agua para
Hemodiálisis, III Jornadas Rioplatenses de Microbiología,
Buenos Aires, Argentinian Association for Microbiology, Oc-
tober 1997; 101.
11. Bambauer R, Schauer M, Jung WK, Daum V, Vienken J. Con-
tamination of dialysis water and dialysate. A survey of 30 cen-
ters. ASAIO J 1994; 40: 1012-6.
Received: January 15, 2002
Revised: April 29, 2002
Accepted: May 30, 2002
© Società Italiana di Nefrologia